Hadamard-encoded sub-slice fMRI for reduced signal dropout
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک ماده چگال
پیش نمایش صفحه اول مقاله
Hadamard-encoded sub-slice fMRI for reduced signal dropout
چکیده انگلیسی
T2⁎-weighted blood oxygen level-dependent functional magnetic resonance imaging is adversely affected by susceptibility-induced field gradients in brain regions adjoining air interfaces that cause image distortion and signal dropout. Reducing slice thickness diminishes signal dropout but is accompanied by reduced signal-to-noise ratio (SNR). This study proposes simultaneous excitation of subslices with total width equal to the desired slice thickness, employing alternating Hadamard-encoded radiofrequency pulses coupled with incoherent addition of the subslices to achieve reduction of through-plane dephasing with minimal SNR loss but at the expense of a reduction in temporal resolution. Using a sensory task and hypercapnic challenge with breathholding (BH), results with two subslices per slice and a twofold reduction in temporal resolution show improved activation relative to a conventional acquisition. Average (eight subjects) T-scores in the BH task increased by 16% (P<.0003), and activation extent increased by 12% (not significant). In frontal brain regions, significant improvements in BH activation extent (11.4%, P<.05) and T-scores (18%, P<.0002) were demonstrated. Higher temporal resolution can be achieved by tradeoff of SNR.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Magnetic Resonance Imaging - Volume 30, Issue 1, January 2012, Pages 1-8