Modified INOvent for delivery of inhaled nitric oxide during cardiac MRI
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک ماده چگال
پیش نمایش صفحه اول مقاله
Modified INOvent for delivery of inhaled nitric oxide during cardiac MRI
چکیده انگلیسی

BackgroundThe aim of this study was to assess the feasibility of delivering NO through a modified system to allow clearance of the magnetic field and thus compatibility with cardiac magnetic resonance (CMR). Nitric oxide (NO) is an inhalational, selective pulmonary vasodilator with a wide range of applications in a variety of disease states, including diseases that affect the right ventricle. Accurate assessment of dynamic changes in right ventricular function necessitates CMR; however, delivery of NO is only possible using equipment that is not magnetic resonance imaging (MRI) compatible (INOvent delivery system, Ohmeda, Inc., Madison, WI, USA).MethodsThe INOvent delivery system was modified by using 35 ft. of standard oxygen tubing to allow NO delivery through an electrical conduit and into the MRI suite. The concentrations of oxygen (O2), nitrogen dioxide (a harmful byproduct, NO2) and NO were measured in triplicate using the built-in electrochemical analyzer on the INOvent. After confirmation of safety, the system was used to administer drug to a patient x, and dynamic MRI measurements were performed.ResultsWhen the standard INOvent was set to administer 40 ppm of NO, the mean/standard deviation of gas delivered was as follows: NO: 42/0 ppm; NO2: 0.3/0.1 ppm; and O2: 93/0 ppm. In comparison, the gas delivery of the modified INOvent was follows: NO: 41/0 ppm; NO2: 0.5/0 ppm; and O2: 93.7/0.6 ppm. During administration to an index patient with severe pulmonic insufficiency (PI), a measurable reduction in PI was observed by CMR.ConclusionsNitric oxide can be administered through 35 ft. of standard oxygen tubing without significantly affecting dose delivery. This technique has potential application in patients with right-sided structural heart disease for determination of dynamic physiological changes.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Magnetic Resonance Imaging - Volume 29, Issue 8, October 2011, Pages 1145–1149
نویسندگان
, , , , , ,