“MR conditional” respiratory ventilator system incident in a 3-T MRI environment
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم فیزیک ماده چگال
پیش نمایش صفحه اول مقاله
“MR conditional” respiratory ventilator system incident in a 3-T MRI environment
چکیده انگلیسی

The misunderstanding of the labeling for an “MR conditional” respiratory ventilator system resulted in the placement of this device, which had substantial ferromagnetic components, too close to a 3-T magnetic resonance (MR) system, causing a projectile incident. Magnetic resonance imaging health care professionals should be aware of the potentially dangerous consequences when such medical devices that have Food and Drug Administration-approved MR conditional labeling are brought into the MR system room. Recommendations to prevent future incidents are provided herein.