• We developed a new micro-CT scanner for small animal imaging.
• Application of Timepix technology to obtain enhanced soft tissue contrast.
• Spatial resolution below 30 µm achieved.
• Performance demonstrated using a tissue equivalent phantom and biological samples.
We describe a newly developed compact micro-CT scanner with rotating gantry equipped with a Timepix Quad hybrid pixel semiconductor detector and a micro-focus X-ray tube providing spatial resolution down to 30 µm. The resolving power of the device in relation to soft tissue sensitivity is demonstrated using a tissue-equivalent phantom and different types of biological samples. The results demonstrate that the use of noiseless particle counting detectors is a promising way to achieve sufficient soft tissue contrast even without any contrast agents.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 773, 11 February 2015, Pages 81–86