کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1823003 1526411 2013 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Memory effect, resolution, and efficiency measurements of an Al2O3 coated plastic scintillator used for radioxenon detection
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Memory effect, resolution, and efficiency measurements of an Al2O3 coated plastic scintillator used for radioxenon detection
چکیده انگلیسی

A cylindrical plastic scintillator cell, used for radioxenon monitoring within the verification regime of the Comprehensive Nuclear-Test-Ban Treaty, has been coated with 425 nm Al2O3 using low temperature Atomic Layer Deposition, and its performance has been evaluated. The motivation is to reduce the memory effect caused by radioxenon diffusing into the plastic scintillator material during measurements, resulting in an elevated detection limit. Measurements with the coated detector show both energy resolution and efficiency comparable to uncoated detectors, and a memory effect reduction of a factor of 1000. Provided that the quality of the detector is maintained for a longer period of time, Al2O3 coatings are believed to be a viable solution to the memory effect problem in question.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 714, 21 June 2013, Pages 128–135
نویسندگان
, , , , , , ,