LET spectrometry of 14 MeV (D–T) neutrons using CR-39 track detectors
چکیده انگلیسی

Linear energy transfer (LET) spectrum in water in the range of 12 keV/μm to 382 keV/μm due to 14 MeV (D–T) neutrons is estimated using the track size parameters in two different types of CR-39 track detectors, viz. Intercast (1.5 mm) and Pershore (0.5 mm). Another set of CR-39s (Intercast) combined with 1 mm polyethylene (PE) radiators is exposed to study the effect of enhanced recoils on the LET spectrum. The detection efficiencies for all these cases and the enhancement ratio due to PE radiator are determined. Using this LET spectrum, the microdosimetric spectra of absorbed doses and dose equivalents are estimated based on the Q–L conversion factors as given in ICRP 60. The shape of the LET spectra are found to be similar in all the cases, however, the dose equivalents obtained with the CR-39+PE radiator is about 20% more than the other detectors without PE. The ratios of dose equivalents obtained from LET spectra (HLET) and the ambient dose equivalent (H*(10)) obtained from fluence-to-dose equivalent conversion factors (ICRP 74) for 14 MeV neutrons are used to estimate the dose response of the detectors. H*(10) is also measured using a neutron rem meter, the response of which is found to be about 23% less than the actual dose.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 708, 21 April 2013, Pages 46–50
نویسندگان
, , , ,