To satisfy the requirements of the next generation of dark matter detectors based on the dual phase TPC, Hamamatsu, in close collaboration with UCLA, has developed the R11410-10 photomultiplier tube. In this work, we present the detailed tests performed on this device. High QE (∼30%∼30%) accompanied by a low dark count rate (50 Hz at 0.3 PE) and high gain (1×107) with good single PE resolution have been observed. A comprehensive screening measurement campaign is ongoing while the manufacturer quotes a radioactivity of 20 mBq/PMT. These characteristics show the R11410-10 to be particularly suitable for the forthcoming zero background liquid xenon detectors.
► Requirements needed of PMTs for dark matter searches.
► Systematic tests of 3-in. Hamamatsu R11410-PMT for dark matter detectors.
► Results show high quantum efficiency, low radioactivity, and low noise properties.
► PMT demonstrates suitable qualities necessary for rare event searches.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 696, 22 December 2012, Pages 32–39