کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1823992 1645430 2012 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Study of the nucleation in superheated emulsion detector at high pressure
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Study of the nucleation in superheated emulsion detector at high pressure
چکیده انگلیسی

The presence of second metastable state in the superheated emulsion detector has been reported earlier. The nucleation of the superheated droplets in a radiation field shows a proportional nature when the droplet depletion is negligible. However, the second metastable state has a much higher nucleation frequency and its population replenishes from the normal metastable state when there is a break in irradiation, and this causes the nucleation of the droplets not to be proportional to the dose delivered. It is observed that the droplets in the second metastable state decreases with increased temperature. The study of nucleation at high temperature and pressure, where the population of the droplets in the second metastable state is expected to be small, is presented here.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 682, 1 August 2012, Pages 31–35