کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1824337 1027333 2011 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Development of a neutron tomography system using a low flux reactor
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Development of a neutron tomography system using a low flux reactor
چکیده انگلیسی
A neutron tomography instrument was designed and developed at the Royal Military College (RMC) of Canada with Queen's University to enhance these institutions' non-destructive evaluation capabilities. The neutron imaging system was built around a Safe Low-Power C(K)ritical Experiment (SLOWPOKE-2) nuclear research reactor. The low power and physical geometry of the reactor required that a novel design be developed to facilitate tomography. A unique rotisserie style rotary stage and clamping apparatus was developed. Furthermore, the low flux at the image plane (3×104 n cm−2 s−1), necessitated that the image acquisition and reconstruction processes be optimized. Tomographs of numerous samples were obtained using the new tomography instrument at RMC.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 651, Issue 1, 21 September 2011, Pages 81-85
نویسندگان
, , , , ,