کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1824646 1027342 2011 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Advances in silicon carbide X-ray detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Advances in silicon carbide X-ray detectors
چکیده انگلیسی

The latest advances in SiC X-ray detectors are presented: a pixel detector coupled to a custom ultra low noise CMOS preamplifier has been characterized at room and high temperature. An equivalent noise energy (ENE) of 113 eV FWHM, corresponding to 6.1 electrons r.m.s., has been achieved with the detector/front-end system operating at +30 °C. A Fano factor of F=0.10 has been estimated from the 55Fe spectrum. When the system is heated up to +100 °C, the measured ENE is 163 eV FWHM (8.9 electrons r.m.s.). It is determined that both at room and at high temperature the performance are fully limited by the noise of the front-end electronics. It is also presented the capability of SiC detectors to operate in environments under unstable temperature conditions without any apparatus for temperature stabilization; it has been proved that a SiC detector can acquire high resolution X-ray spectra without spectral line degradation while the system temperature changes between +30 and +75 °C.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 652, Issue 1, 1 October 2011, Pages 193–196
نویسندگان
, , , ,