CZT performance for different anode pixel geometries and data corrections
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
CZT performance for different anode pixel geometries and data corrections
چکیده انگلیسی

We have investigated the energy resolution and absolute photopeak efficiency of CZT detectors with three anode pixel geometries after three data correction steps. Before data corrections, the energy resolution is 4–5% at 122 keV, and the efficiency is 46–70% for a 21 keV photopeak window. Charge sharing events, which make up 10–25% of all events, are a significant cause of the poor efficiency. We discovered that large gaps between anode pixels have a strong negative impact on both charge sharing and absolute photopeak efficiency. Using three-dimensional position-dependent energy corrections, we were able to compensate for some of this reduced sensitivity, increasing the efficiency by up to 35% to a final efficiency of 62–80% of incoming gamma events.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 648, Supplement 1, 21 August 2011, Pages S37–S41