کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1825094 1027353 2011 10 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A novel γ‐rayγ‐ray imaging method for the pulse-shape characterization of position sensitive semiconductor radiation detectors
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
A novel γ‐rayγ‐ray imaging method for the pulse-shape characterization of position sensitive semiconductor radiation detectors
چکیده انگلیسی

A new technique for the pulse-shape characterization of γ‐rayγ‐ray position sensitive germanium detectors is presented. This method combines the pulse shape comparison scan (PSCS) principle with a γ‐rayγ‐ray imaging technique. The latter is provided by a supplementary, high performance, position sensitive γ‐rayγ‐ray scintillator detector. We describe the basic aspects of the method and we show measurements made for the study of pulse-shapes in a non-segmented planar HPGe detector. A preliminary application of the PSCS is carried out, although a more detailed investigation is being performed with highly segmented position sensitive detectors.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 643, Issue 1, 1 July 2011, Pages 79–88
نویسندگان
, , , , , , , , ,