کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1825126 1526463 2011 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Experimental response functions of a single-crystal diamond detector for 5–20.5 MeV neutrons
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Experimental response functions of a single-crystal diamond detector for 5–20.5 MeV neutrons
چکیده انگلیسی

A single crystal diamond detector was exposed to quasi mono-energetic neutron fields in the energy range from 5 to 20.5 MeV. Response functions of such a detector were measured for the first time for neutron energies above 14 MeV. The pulse height spectra showed sharp peaks at specific energies. These peaks result from neutron induced charged particle reactions occurring in carbon at different neutron energies. The centroid of the peaks due to the 12C(n,α)9Be reaction permitted, by means of the reaction Q-values, to point out a good linearity of the detector response versus neutron energy and a good intrinsic energy resolution of about 56 keV at full width at half maximum. The results reveal that single crystal diamond detectors are good candidates as high resolution fast neutron spectrometers for nuclear fusion applications and for systematic studies of beam and target characteristics for neutron-producing reactions.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 640, Issue 1, 1 June 2011, Pages 185–191
نویسندگان
, , , , , ,