Construction and tests of demonstrator modules for a 3-D axial PET system for brain or small animal imaging
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Construction and tests of demonstrator modules for a 3-D axial PET system for brain or small animal imaging
چکیده انگلیسی

The design and construction of a PET camera module with high sensitivity, full 3-D spatial reconstruction and very good energy resolution is presented. The basic principle consists of an axial arrangement of long scintillation crystals around the Field Of View (FOV), providing a measurement of the transverse coordinates of the interacting 511 keV gamma ray. On top of each layer of crystals, an array of Wave-Length Shifter (WLS) strips, which collect the light leaving the crystals sideways, is positioned orthogonal to the crystal direction. The signals in the WLS strips allow a precise measurement of the z (axial) co-ordinate of the 511 keV γ-ray gamma impact. The construction of two modules used for demonstration of the concept is described. First preliminary results on spatial and energy resolution from one full module will be shown.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 636, Issue 1, Supplement, 21 April 2011, Pages S226–S230