کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1825590 1027365 2011 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Demonstration of an Axial PET concept for brain and small animal imaging
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Demonstration of an Axial PET concept for brain and small animal imaging
چکیده انگلیسی

Standard Positron Emission Tomography (PET) cameras need to reach a compromise between spatial resolution and sensitivity. To overcome this limitation we developed a novel concept of PET. Our AX-PET demonstrator is made of LYSO crystals aligned along the z coordinate (patient's axis) and WLS strips orthogonally placed with respect to the crystals. This concept offers full 3D localization of the photon interaction inside the camera. Thus the spatial resolution and the sensitivity can be simultaneously improved and the reconstruction of Compton interactions inside the detector is also possible. Moreover, by means of G-APDs for reading out the photons, both from LYSO and WLS, the detector is insensitive to magnetic fields and it is then suitable to be used in a combined PET/MRI apparatus. A complete Monte Carlo simulation and dedicated reconstruction software have been developed. The two final modules, each composed of 48 crystals and 156 WLS strips, have been built and fully characterized in a dedicated test set-up. The results, obtained with a 22Na point source (0.25 mm diameter), of the single module performances and a first estimation of the performances with the two module system are reported.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 628, Issue 1, 1 February 2011, Pages 426–429
نویسندگان
, , , , , , , , , , , , , , , , , , , , ,