کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1825982 | 1027373 | 2010 | 7 صفحه PDF | دانلود رایگان |

The GAINS (Germanium Array for Inelastic Neutron Scattering) setup was recently installed at the GELINA (Geel Linear Accelerator) neutron source of EC-JRC-IRMM. In order to produce reliable neutron inelastic cross-section data the setup requires a precise efficiency calibration for each sample under investigation. These samples represent extended sources in non-standard orientations and with significant self-attenuation of gamma rays. We present in detail the efficiency calibration procedure performed with two Monte Carlo simulations using MCNP and GEANT4. A good agreement is found between the two independent approaches. The general features of the simulations are discussed together with an in-depth description of the various factors influencing the efficiency.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 624, Issue 1, 1 December 2010, Pages 130–136