کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1826546 1027385 2011 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Performance study of a fan beam collimator designed for a multi-modality small animal imaging device
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Performance study of a fan beam collimator designed for a multi-modality small animal imaging device
چکیده انگلیسی

This paper describes the methodology to design and conduct the performances of a fan beam collimator. This fan beam collimator was designed to use with a multi-modality small animal imaging device and the performance of the collimator was studied for a 3D geometry. Analytical expressions were formulated to calculate the parameters for the collimator. A Monte Carlo model was developed to analyze the scattering and image noises for a 3D object. The results showed that the performance of the fan beam collimator was strongly dependent on the source distribution and position. The fan beam collimator showed increased counting efficiency in comparison to a parallel hole collimator. Inside attenuating medium, the increased attenuating effect outweighed the fan beam increased counting efficiency.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 629, Issue 1, 11 February 2011, Pages 368–376
نویسندگان
, , , ,