کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1826603 1027387 2010 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Design and performance of a Compton-coincidence system for measuring non-proportionality of new scintillators
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Design and performance of a Compton-coincidence system for measuring non-proportionality of new scintillators
چکیده انگلیسی

A Compton-coincidence scintillator light yield non-proportionality measurement system was constructed and tested. Improved testing procedures and timing resolution have allowed the use of significantly less active excitation sources without lengthening test times or degrading data integrity. Previous methods typically required 100 mCi Cs-137 sources, but an increase in solid angle between source and sample as well as improved timing resolution have allowed the use of sources with activities less than 10 μCi. Representative samples of BGO and NaI(Tl) were tested to validate the system against literature single-source data.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 615, Issue 2, 1 April 2010, Pages 182–187
نویسندگان
, , ,