کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1826961 1027400 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Analysis of SMELS and reference materials for validation of the k0-based internal monostandard NAA method using in-situ detection efficiency
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Analysis of SMELS and reference materials for validation of the k0-based internal monostandard NAA method using in-situ detection efficiency
چکیده انگلیسی

Three synthetic multielement standards (SMELS I, II and III) and two reference materials (RMs), SL-3 and Soil-7 of IAEA were analyzed for validation of the k0-based internal monostandard neutron activation analysis (IM-NAA) method utilizing in-situ relative detection efficiency. The internal monostandards used in SMELS and RMs were Au and Sc, respectively. The samples were irradiated in Apsara and Dhruva reactors, BARC and radioactive assay was carried out using a 40% relative efficiency HPGe detector coupled to an 8 k MCA. Concentrations of 23 elements were determined in both SMELS and RMs. In the case of RMs, concentrations of a few elements, whose certified values are not available, could also be determined. The % deviations for the elements determined in SMELS with respect to the assigned values and RMs with respect to certified values were within ±8%. The Z-score values at 95% confidence level for most of the elements in both the materials were within ±1.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 622, Issue 2, 11 October 2010, Pages 411–414
نویسندگان
, , ,