Calculation of responses and analysis of experimental data for a silicon gamma spectrometer
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
ابزار دقیق
پیش نمایش صفحه اول مقاله
چکیده انگلیسی
A silicon detector developed for gamma ray spectrometry (intended for measurement in the mixed neutron gamma fields, e.g. in reactor physics) was tested for the photon energy interval up to 10 MeV. The responses of this spectrometric system were simulated using a Monte Carlo method (MCNPX code), and the response matrix was calculated. Unfolding techniques for reconstructing incident photon spectra were tested, and a method based on the Scofield–Gold iterative solution of the corresponding matrix equation was applied. The applicability of such an approach to spectrometry data processing and spectral analysis is demonstrated, an example of the results is presented, and possible future improvements and optimizations are discussed.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 619, Issues 1–3, 1–21 July 2010, Pages 186–189
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 619, Issues 1–3, 1–21 July 2010, Pages 186–189
نویسندگان
Jaroslav Klusoň, Bohumil Jánský,