A novel design for an inexpensive depth of interaction capable detector for γγ-ray imaging has been developed. The design takes advantage of the strong correlation between the width of the scintillation light distribution in monolithic crystals and the interaction depth of γγ-rays. We present in this work an inexpensive modification of the commonly used charge dividing circuits which enables the instantaneous and simultaneous computation of the second order moment of light distribution. This measure provides a good estimate for the depth of interaction and does not affect the determination of the position centroids and the energy release of γγ-ray impact. The method has been tested with a detector consisting of a monolithic LSO block sized 42×42×10mm3 and a position-sensitive photomultiplier tube H8500 from Hamamatsu. The mean spatial resolution of the detector was found to be 3.4mm for the position centroids and 4.9mm for the DOI. The best spatial resolutions were observed at the center of the detector and yielded 1.4mm for the position centroids and 1.9mm for the DOI.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 600, Issue 3, 11 March 2009, Pages 624–634