کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1828589 1526483 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Digital pulse shape analysis for the capture-gated liquid scintillator BC-523A
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Digital pulse shape analysis for the capture-gated liquid scintillator BC-523A
چکیده انگلیسی

The BC-523A detector is a boron-10-loaded liquid scintillation detector that is sensitive to not only fast neutrons and gamma rays, but also to thermal neutrons. This detector has recently attracted the attention of several researchers and has shown promise for neutron spectroscopy. Several efforts have been made to accurately characterize this detector for several neutron and gamma-ray energies. In this paper, we show, for the first time, that neutron capture pulses can be visually distinguished from neutron scattering pulses by using an offline digital pulse shape discrimination method based on charge integration. In addition, new Monte Carlo simulation and measurement results for time to capture are compared and reasonable agreement is achieved. This work has applications in fields such as nuclear safeguards, nuclear nonproliferation, and national security.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 599, Issues 2–3, 11 February 2009, Pages 221–225
نویسندگان
, ,