کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1829380 1027457 2008 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Light collection optimization in scintillator-based gamma-ray spectrometers
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Light collection optimization in scintillator-based gamma-ray spectrometers
چکیده انگلیسی

Scintillator-based gamma-ray detectors are being actively pursued for homeland security applications. A key property of such detectors is their energy resolution which enables faster detection and more precise identification of gamma-ray sources. In order to obtain the best energy resolution with a given scintillator material, it is crucial to collect the largest fraction possible of the light emitted after gamma-ray absorption. Different techniques to maximize the light collection efficiency were investigated and tested experimentally. In particular, the effect of the scintillator geometry has been simulated with Detect2000. Also, a number of wrapping materials have been tested for their reflectivity and their performance in terms of improving the energy resolution in a BGO-based gamma-ray detector. The best results were obtained with a tapered cylinder geometry and the GORE DRP tape.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 588, Issue 3, 11 April 2008, Pages 384–388