A new fast two-stage GL charge division method for a very long MSGC for large-area radiation imaging
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
A new fast two-stage GL charge division method for a very long MSGC for large-area radiation imaging
چکیده انگلیسی
We present a novel readout method; a fast two-stage Global Local (GL) charge division method. The new readout method is to utilize two coincident signals to localize position. It can greatly improve a spatial resolution when applies to a high signal-to-noise ratio (SNR) detector. We have implemented the new readout method in very long microstrip gas counters (MSGCs). Prototype detectors, which have 640.8 mm of an active length, have been designed, fabricated and evaluated. The principle of the new readout method was confirmed. The spatial resolution was 1.15 mm with good uniformity and good linearity.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 580, Issue 2, 1 October 2007, Pages 1119-1122