کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1830288 1027476 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Ion chamber monitors for pulsed X-ray background at electron accelerator facilities
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Ion chamber monitors for pulsed X-ray background at electron accelerator facilities
چکیده انگلیسی

Ion chambers sensitive to pulsed X-rays have been developed with special electrode configuration of four electrodes to suit the area-monitoring requirements at the electron accelerators. For comparison, a chamber with similar sensitive volume and overall dimensions was developed with a standard electrode configuration. The standard chamber can measure gamma radiation from 100 μR/h to 1 R/h range whereas the four-electrode chamber can measure gamma radiation from 100 μR/h to 1 kR/h with ±10% signal linearity thus enhancing three decades of measuring range. The chambers have been tested in pulsed X-ray background at the INDUS-I facility at R.R. Centre for Advanced Technology, Indore, under normal as well as beam loss conditions and the four-electrode chamber shows superior performance compared to a standard chamber. The experimental results obtained compared well with the theoretically estimated values.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 578, Issue 1, 21 July 2007, Pages 246–252
نویسندگان
, ,