کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1830443 1526496 2007 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators
چکیده انگلیسی

The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by −0.11° compared with the digital indicator, and the standard deviation was within 0.05°. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 576, Issues 2–3, 21 June 2007, Pages 441–445
نویسندگان
, , , , ,