کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1830988 1526498 2007 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Signal analysis and processing for SmartPET
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم ابزار دقیق
پیش نمایش صفحه اول مقاله
Signal analysis and processing for SmartPET
چکیده انگلیسی

Measurement of induced transient charges on spectator electrodes is a critical requirement of the SmartPET project. Such a task requires the precise measurement of small amplitude pulses. Induced charge magnitudes on the SmartPET detectors were therefore studied and the suitability of wavelet analysis applied to de-noising signals was investigated. It was found that the absolute net maximum induced charge magnitudes from the two adjacent electrodes to the collecting electrode is 17% of the real charge magnitude for the AC side and 20% for the DC side. It was also found that wavelet analysis could identify induced charges of comparable magnitude to system noise.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 573, Issues 1–2, 1 April 2007, Pages 95–98
نویسندگان
, , , , , , , ,