کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1831641 | 1027500 | 2007 | 8 صفحه PDF | دانلود رایگان |

A high spatial resolution neutron imaging device was developed at the Mark II TRIGA reactor at The University of Texas at Austin. As the modulation transfer function (MTF) is recognized as a well-established parameter for evaluation of imaging system resolution, the aliasing associated with digital sampling adds complexity to its measurement. Aliasing is especially problematic when using a high spatial resolution micro-channel plate (MCP) neutron detector that has a pixel grid size similar to that of a CCD array. To compensate for the aliasing an angulated edge method was used to evaluate the neutron imaging facility, overcoming aliasing by obtaining an oversampled edge spread function (ESF). Baseline correction was applied to the ESF to remove the noticeable trends and the LSF was multiplied by Hann window to obtain a smoothed version of presampled MTF. The computing procedure is confirmed by visual inspection of a testing phantom; in addition, it is confirmed by comparison to the MTF measurement of a scintillation screen with a known MTF curve.
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment - Volume 582, Issue 2, 21 November 2007, Pages 621–628