کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2161300 1090911 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Notes on the construction of solid-state detectors
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی تحقیقات سرطان
پیش نمایش صفحه اول مقاله
Notes on the construction of solid-state detectors
چکیده انگلیسی

Background and purposeSoild-state detectors such as diodes and diamonds are useful radiation detectors, particularly for small field measurements, such as in stereotactic dosimetry [6] or in the small sub-field measurements required for intensity modulated radiotherapy (IMRT) [5]. In all small field measurements, positioning of the detector is critical. Generally it is assumed that the center of the sensitive volume of each solid-state detector is coincident with the geometric center of the outer casing.Patients and methodsX-rays were taken of four detectors (diamond, three types of diode) to determine the physical characteristics of each, particularly the position of the centre of the sensitive volume.ResultsThe results showed that the position of the centre of the sensitive volume is not always coincident with the geometric centre of a soild-state detector.ConclusionsThe construction of a detector used for precise measurements should be investigated prior to use, by taking high resolution x-ray images.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Radiotherapy and Oncology - Volume 79, Issue 3, June 2006, Pages 348–351
نویسندگان
, ,