کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2464984 1111814 2010 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Measurement of platelet function in dogs using a novel impedance aggregometer
موضوعات مرتبط
علوم زیستی و بیوفناوری علوم کشاورزی و بیولوژیک علوم دامی و جانورشناسی
پیش نمایش صفحه اول مقاله
Measurement of platelet function in dogs using a novel impedance aggregometer
چکیده انگلیسی

The aim of this study was to optimise the technique and establish reference values for whole blood aggregometry in dogs using a novel multiplate analyser. Measurements were performed on the hirudin-anticoagulated blood of healthy dogs using a wide range of agonists. Optimal agonist concentrations were 10 μmol/L of adenosine diphosphate, 5 μg/mL of collagen and 1 mmol/L of arachidonic acid. Ristocetin (at 0.2 and 1 mg/mL) and thrombin receptor activating peptide (TRAP-6 at 32 and 160 μmol/L) did not consistently induce platelet aggregation. Coefficients of variance for within-run imprecision (n = 10 repetitions) varied from 5% to 18%. Measurement signals were significantly higher when analyses were performed on standard samples (hirudin-anticoagulated blood) compared to citrated blood or blood samples anticoagulated with citrate buffer, regardless of whether or not re-calcification was performed (P < 0.05). The findings indicate that the analyser is suitable for the investigation of platelet aggregation in dogs and analysis should be performed on hirudin-anticoagulated blood using optimised agonist concentrations.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Veterinary Journal - Volume 185, Issue 2, August 2010, Pages 144–151
نویسندگان
, , ,