کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
2691264 | 1143332 | 2009 | 7 صفحه PDF | دانلود رایگان |

Study DesignClinical measurement, cross sectional.PurposeTo introduce a new measurement device, the Pollexograph®, to easily measure palmar thumb abduction, and to compare its reliability with conventional goniometry.MethodsFourteen hand therapists measured palmar abduction of the same healthy subject with the Pollexograph and a conventional goniometer. In addition, intrarater reliability of the Pollexograph was studied in 21 patients with a hypoplastic thumb.ResultsVariance between measurements of the same subject measured by the hand therapist was 2–6 times smaller with the Pollexograph compared to conventional goniometry. Pollexograph intrarater reliability in hypoplastic thumb patients was excellent (intraclass correlation coefficient (ICC) = 0.98–0.99).ConclusionsA new tool to measure palmar abduction in clinical care, the Pollexograph, has been introduced. The Pollexograph reduces variability between raters when measuring the same subject compared with conventional goniometry and excellent measurement reliability in hypoplastic thumb patients.Level of evidenceNot applicable.
Journal: Journal of Hand Therapy - Volume 22, Issue 3, July–September 2009, Pages 271–277