Reliability of the Skin Compliance Device in the Assessment of Scar Pliability
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
علوم پزشکی و سلامت
پزشکی و دندانپزشکی
ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
The purpose of the study was to determine the intrarater and interrater reliability of the Skin Compliance Device, a tool to measure scar pliability. Three points on 25 scars and the corresponding areas on the uninvolved upper extremities were tested by two evaluators. Testing locations were standardized using a wound-tracing device. Intrarater reliability was excellent for both raters for all points tested on both extremities (intraclass correlation coefficient [ICC]Â >Â 0.88). On the involved side, good inter-rater reliability was found for the most adherent point (ICCÂ =Â 0.79) and the most distal point (ICCÂ =Â 0.77). Fair interrater reliability was found for the most proximal point on the involved side (ICCÂ =Â 0.29) and fair-to-moderate reliability for all points on the unscarred extremity. The Skin Compliance Device demonstrates excellent intrarater reliability and good interrater reliability for measuring the most adherent aspect of scars.