کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2693524 1143546 2015 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of corneal deformation analyzed with a Scheimpflug based device
ترجمه فارسی عنوان
ارزیابی تغییر شکل قرنیه با استفاده از دستگاه مبتنی بر Schemapflug تجزیه و تحلیل شده
کلمات کلیدی
بیومکانیک قرنل؛ پچیمتری قرنیه؛ چشمان ناآشکار؛ ارزیابی توموگرافی؛ ارزیابی بیومکانیک
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
چکیده انگلیسی


• Corneal deformation has been measured in healthy eyes with Corvis ST.
• Corneal curvature and corneal thickness have been measured in the same group.
• Correlation between deformation and morphological parameters have been run.
• Corneal deformation measured by Corvis ST seem to be indipendent by morphological parameters.

PurposeTo evaluate the correlation between corneal biomechanical and morphological data in healthy eyes.MethodsA complete clinical eye examination of naïve eyes was followed by tomographic (Pentacam, Oculus, Wetzlar, Germany) and biomechanical (Corvis ST, Oculus, Wetzlar, Germany) evaluation. Linear regression between central corneal thickness (CCT), corneal volume (CV) and anterior corneal curvature measured with Sim’K (SK), versus corneal deformation parameters measured with Corvis ST have been run using SPSS software version 18.0.ResultsSeventy-six eyes of 76 healthy subjects (44 women and 32 men) with a mean age of 36.84 ± 10.74 years and a mean refractive error of −0.55 ± 1.68 D (measured as spherical equivalent) were evaluated. Corneal deformation parameters were weakly correlated with corneal morphological parameters and with spherical equivalent. Although the correlations between deformation amplitude versus SK and between SK versus Velocity of Applanation 2, were higher than the others (R2 = 0.28 and 0.26 respectively), none of them was statistically significant (p > 0.01).ConclusionsAccording with these findings, Corvis ST seems to be able to provide an analysis of corneal deformation independent from corneal morphological characteristics. If these data will be confirmed in further studies, this device could be useful in the management and screening of eyes with corneal diseases.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 38, Issue 2, April 2015, Pages 89–93
نویسندگان
, , , , , , ,