Peripheral refraction with eye and head rotation with contact lenses
ترجمه فارسی عنوان
انکسار محیطی با چرخش چشم و سر با لنزهای تماسی
کلمات کلیدی
انکسار محیطی؛ نزدیک بینی؛ تماس با لنز چند کانونی؛ چرخش چشم؛ چرخش سر
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
چکیده انگلیسی


• We evaluated the effect of eye and head rotation in peripheral refraction with contact lenses.
• We found no differences between measurements obtained by eye and head rotation.
• Multifocal center-distance soft contact lenses induced peripheral myopization.
• Multifocal concentric soft contact lenses did not show changes in peripheral refraction.

PurposeTo evaluate the impact of eye and head rotation in the measurement of peripheral refraction with an open-field autorefractometer in myopic eyes wearing two different center-distance designs of multifocal contact lenses (MFCLs).MethodsNineteen right eyes from 19 myopic patients (average central M ± SD = −2.67 ± 1.66 D) aged 20–27 years (mean ± SD = 23.2 ± 3.3 years) were evaluated using a Grand-Seiko autorefractometer. Patients were fitted with one multifocal aspheric center-distance contact lens (Biofinity Multifocal D®) and with one multi-concentric MFCL (Acuvue Oasys for Presbyopia). Axial and peripheral refraction were evaluated by eye rotation and by head rotation under naked eye condition and with each MFCL fitted randomly and in independent sessions.ResultsFor the naked eye, refractive pattern (M, J0 and J45) across the central 60° of the horizontal visual field values did not show significant changes measured by rotating the eye or rotating the head (p > 0.05). Similar results were obtained wearing the Biofinity D, for both testing methods, no obtaining significant differences to M, J0 and J45 values (p > 0.05). For Acuvue Oasys for presbyopia, also no differences were found when comparing measurements obtained by eye and head rotation (p > 0.05). Multivariate analysis did not showed a significant interaction between testing method and lens type neither with measuring locations (MANOVA, p > 0.05). There were significant differences in M and J0 values between naked eyes and each MFCL.ConclusionMeasurements of peripheral refraction by rotating the eye or rotating the head in myopic patients wearing dominant design or multi-concentric multifocal silicone hydrogel contact lens are comparable.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 38, Issue 2, April 2015, Pages 104–109
نویسندگان
, , , , , ,