Accuracy and sensitivity of the dynamic ocular thermography and inter-subjects ocular surface temperature (OST) in Chinese young adults
چکیده انگلیسی

PurposeTo investigate the accuracy and sensitivity of the Thermo-Tracer TH9100MV, NEC San-ei in monitoring the temperature of the anterior eye surface and to establish an inter-subjects ocular surface temperature (OST) norm in Chinese young adults.MethodsA blackbody with fixed temperature was used to study accuracy and sensitivity of the thermal camera used (Thermo-Tracer TH9100MV, NEC San-ei). Sixty healthy Chinese young adults aged 17–21 (mean age 19) were recruited from the Singapore Polytechnic Optometry Centre. A room with standard room temperature and humidity was used. Subjects were adapted to the room conditions for 20 min before measurement. OST was then measured within 4–5 s after a blink and with their upper lid lifted by a cotton bud, using infrared thermo-camera, in a non-invasive manner.ResultsAll future Thermo-Tracer readings could be converted to ‘true’ temperature via the regression equation derived y = 1.0884x − 3.4026 (r2 = 0.9985), where y is the ‘true’ temperature and x is reading by Thermo-Tracer. Sensitivity of the machine for all 16 temperatures studied was found to be 91.3%. Mean OST for the population studied was 34.70 ± 0.50 °C. Mean OST at geometrical centre cornea (GCC) was 34.39 ± 0.47 °C and inter-ocular temperature difference was ∼0.20 ± 0.15 °C. Both the horizontal and vertical OST profile was established. No significant differences between right/left eyes and between genders were found at 95% CI.ConclusionsFindings on the accuracy can be used as a reference by researchers who are using the same infrared thermal camera. It is proven that the machine has high sensitivity in determining the temperature of anterior ocular surface. Methodology used in this study can also be applied to the determination of the accuracy and sensitivity of the dynamic range for other infrared thermal cameras. Inter-subjects OST norm in Chinese young adults aged 17–21 was established.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 32, Issue 2, April 2009, Pages 78–83
نویسندگان
, , ,