عنوان انگلیسی مقاله ISI
Agreement and reliability in measuring central corneal thickness with a rotating Scheimpflug–Placido system and ultrasound pachymetry
ترجمه فارسی عنوان
توافقنامه و قابلیت اطمینان در اندازه گیری ضخامت قرنیه مرکزی با سیستم Scheimpflug-Placido چرخشی و سنجش اولتراسوند
کلمات کلیدی
توموگرافی قرنیه Scheimpflug؛ پاکی متری اولتراسوند؛ ضخامت قرنیه مرکزی؛ توافق؛ قابلیت اطمینان
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی چشم پزشکی
چکیده انگلیسی

PurposeWe compare the agreement and the reliability in measuring central corneal thickness (CCT) using two different technologies.MethodThe right eyes of 35 healthy individuals who had a negative history of ophthalmic disease, or ocular surgery were examined. The CCT was determined sequentially with a rotating Scheimpflug camera (Sirius; CSO), and an ultrasound pachymeter (P-1; Takagi). For statistical analysis, we used the methods suggested by Bland and Altman.ResultsThe mean values of CCT obtained from Sirius, and ultrasound were 537 ± 28 μm, and 550 ± 35 μm, respectively. There was a high correlation between Sirius and ultrasound (r = 0.92; p < 0.001), but the difference between the two measurements was statistically significant (t = −5.7; p < 0.00001). The precision of Sirius and ultrasound were 9.4 and 15.9 μm; repeatability 13.3 and 22.4 μm, and coefficient of variation 0.9% and 1.5%, respectively. The intraclass correlation coefficient was 0.97 for Sirius and 0.95 for ultrasound.ConclusionsThe average difference between corneal thickness measured with Sirius and ultrasound pachymetry was small but clinically significant. This means that the two instruments cannot be used interchangeably. Sirius showed precision and repeatability almost twice as much as ultrasound pachymetry. Confidence interval of 13.3 μm for Sirius can show variations in corneal thickness with an uncertainty value lower than 2.5% in 95% of cases. The simplicity of use, the possibility to obtain pachymetric maps, and less invasiveness make this instrument potentially useful in contact lens practice.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Contact Lens and Anterior Eye - Volume 37, Issue 6, December 2014, Pages 442–446
نویسندگان
, , , ,