کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2718234 1145475 2013 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Biomechanical Study: Resistance Comparison of Posterior Antiglide Plate and Lateral Plate on Synthetic Bone Models Simulating Danis-Weber B Malleolar Fractures
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Biomechanical Study: Resistance Comparison of Posterior Antiglide Plate and Lateral Plate on Synthetic Bone Models Simulating Danis-Weber B Malleolar Fractures
چکیده انگلیسی

ObjectiveThe purpose of this study was to compare different positions of plates in lateral malleolar Danis-Weber B fractures on synthetic bone: a lateral plate and a posterior antiglide plate.MethodsShort oblique fractures of distal fibula at the level of the syndesmosys were simulated with a fibular osteotomy in sixteen synthetic fibula bones (Synbone®). Eight fractures were fixed with lateral plating associated with an independent lag screw, and the other eight were fixed with posterior antiglide plating with a lag screw through the plate. A strain gage was installed at the center of each plate at the osteotomy site. Supination and external rotation forces were applied to each of the two groups at the bend.ResultsThe lateral position plate group suffered more deformity in response to supination forces compared to the group with the posterior antiglide plate, but this result was not statistically significant. In the tests with external rotation forces, the posterior antiglide plating group had significantly higher resistance (p < 0.05).ConclusionWhen subjected to external rotation forces, osteosynthesis with posterior antiglide plate models simulating type B fractures of the lateral malleolus of the ankle is more resistant than that of the neutralization plate.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Revista Brasileira de Ortopedia (English Edition) - Volume 48, Issue 3, May–June 2013, Pages 221–227
نویسندگان
, , , , , , ,