Performance of heated wire humidifiers: An in vitro study
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی بیهوشی و پزشکی درد
پیش نمایش صفحه اول مقاله
Performance of heated wire humidifiers: An in vitro study
چکیده انگلیسی

PurposeWe assessed the performance of heated wire humidifiers (HWHs), which should avoid water condensation in the circuit.MethodsWe evaluated the efficiency of 3 HWHs, MR850 (Fisher & Paykel, Auckland, New Zealand), CONCHATHERM IV (Hudson RCI, Temecula, Calif), and DAR HC 2000 (Mallinckrodt DAR, Mirandola, Italy), in comparison with that of the MR730 heated humidifier (HH), which has a standard circuit. We measured gas temperature and absolute humidity (AH) at the Y piece of the ventilatory circuit using a test lung ventilated at 2 minute ventilation volumes (5 and 15 L/min). Temperature levels at the Y piece of the ventilatory circuit of the HHs were set at 35°C, 37°C, and 39°C with different gradients (−2°C, 0°C, and +2°C) between the outlet chamber and the Y piece of the ventilatory circuit.ResultsAt the set temperature levels of 35°C, 37°C, and 39°C with a gradient of 0°C, the MR850 and CONCHATHERM IV had lower gas temperature and AH levels as compared with the DAR HC 2000 and MR730 HH. With increasing temperature gradient, gas temperature increased only with the CONCHATHERM IV but AH increased with all the HWHs. The MR850 showed lower gas temperature and AH levels as compared with CONCHATHERM IV. The condensate was abolished inside the inspiratory circuit with the HWHs.ConclusionsHeated wire humidifiers eliminate water condensation but present significant differences in gas temperature and AH levels that are lower than the expected settings.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Critical Care - Volume 22, Issue 3, September 2007, Pages 258–264
نویسندگان
, , , , , , , , , ,