Instrument Accuracy and Reproducibility in Measurements of Pulmonary Function
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Instrument Accuracy and Reproducibility in Measurements of Pulmonary Function
چکیده انگلیسی

BackgroundThe objective of the study was to quantify the accuracy and reproducibility of five commercially available pulmonary function test (PFT) instruments (Collins CPL [Ferraris Respiratory; Louisville, CO]; Morgan Transflow Test PFT System [Morgan Scientific; Haverhill, MA]; SensorMedics V˙max 22D [VIASYS Healthcare; Yorba Linda, CA]; Jaeger USA Masterscreen Diffusion TP [VIASYS Healthcare]; and Medical Graphics Profiler DX System [Medical Graphics Corp; St. Paul, MN]) that are associated with spirometry and the measurement of pulmonary diffusing capacity.MethodsIn a multifactor, single-center, repeated-measures, full factorial 90-day study, a pulmonary waveform generator and a single-breath simulator of diffusing capacity of the lung for carbon monoxide (Dlco) were used to perform simulations of FVC and Dlco maneuvers. Accuracy was assessed as the difference between the observed and simulated values. Reproducibility was determined as the coefficient of variation of all measurements made during the study.ResultsAll instruments demonstrated a high degree of accuracy in the measurement of FVC and FEV1. Overall, the accuracies associated with the measurement of peak flow, forced expiratory flow, mid-expiratory phase, and diffusing capacity were generally lower and more variable among the instruments tested. The coefficients of variation of Dlco measurements over 90 days were higher than those observed for spirometry.ConclusionsThis study demonstrates the feasibility of assessing the accuracy and reproducibility of modern PFT instruments using simulation testing. Our results provide an assessment of the component of PFT accuracy and reproducibility that is due to instrumentation alone.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Chest - Volume 132, Issue 2, August 2007, Pages 388–395
نویسندگان
, , , , , ,