ObjectiveTo evaluate the proximal and distal (iliac) fixation of seven self-expanding endografts, used in the endovascular treatment (EVAR) of abdominal-aortic aneurysm (AAA), by measuring the displacement force (DF) necessary to dislocate the devices from their fixation sites.MethodsA total of 20 human cadaveric aortas were exposed, left in situ and transected to serve as fixation zones. The Anaconda, EndoFit aorto-uni-iliac, Endurant, Powerlink, Excluder, Talent and Zenith stent grafts were deployed and caudal force was applied at the flow divider, through a force gauge. The DF needed to dislocate each device ≥ 20 mm from the infrarenal neck was recorded before and after moulding-balloon dilatation. Cephalad force was similarly applied to each iliac limb to assess distal fixation before and after moulding-balloon dilatation.ResultsEndografts with fixation hooks or barbs displayed a significantly higher DF necessary to dislocate the proximal portion compared with devices with no such fixation modalities (p < 0.001). Balloon dilatation produced a significant increase in DF in both devices with (p < 0.001) or without (p = 0.003) hooks or barbs. Suprarenal support did not enhance proximal fixation (p = 0.90). Balloon dilatation significantly increased the DF necessary to dislodge the iliac limbs (p = 0.007).ConclusionsDevices with fixation hooks displayed higher proximal fixation. Moulding-balloon dilatation increased proximal and distal fixation. Suprarenal support did not affect proximal fixation.
Journal: European Journal of Vascular and Endovascular Surgery - Volume 40, Issue 4, October 2010, Pages 429–435