کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2927953 1176188 2015 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD
چکیده انگلیسی

Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Indian Heart Journal - Volume 67, Issue 1, January–February 2015, Pages 56–59
نویسندگان
, , , , , ,