Minimally invasive system for baroreflex activation therapy chronically lowers blood pressure with pacemaker-like safety profile: results from the Barostim neo trial
موضوعات مرتبط
علوم زیستی و بیوفناوری علم عصب شناسی سیستم های درون ریز و اتونومیک
پیش نمایش صفحه اول مقاله
Minimally invasive system for baroreflex activation therapy chronically lowers blood pressure with pacemaker-like safety profile: results from the Barostim neo trial
چکیده انگلیسی

BackgroundPrevious trials have demonstrated clinically significant and durable reductions in arterial pressure from baroreflex activation therapy (BAT), resulting from electrical stimulation of the carotid sinus using a novel implantable device. A second-generation system for delivering BAT, the Barostim neo™ system, has been designed to deliver BAT with a simpler device and implant procedure.MethodsBAT, delivered with the advanced system, was evaluated in a single-arm, open-label study of patients with resistant hypertension, defined as resting systolic blood pressure (SBP) ≥140 mm Hg despite treatment with ≥3 medications, including ≥1 diuretic. Stable medical therapy was required for ≥4 weeks before establishing pretreatment baseline by averaging two SBP readings taken ≥24 hours apart.ResultsThirty patients enrolled from seven centers in Europe and Canada. From a baseline of 171.7 ± 20.2/99.5 ± 13.9 mm Hg, arterial pressure decreased by 26.0 ± 4.4/12.4 ± 2.5 mm Hg at 6 months. In a subset (n = 6) of patients with prior renal nerve ablation, arterial pressure decreased by 22.3 ± 9.8 mm Hg. Background medical therapy for hypertension was unchanged during follow-up. Three minor procedure-related complications occurred within 30 days of implant. All complications resolved without sequelae.ConclusionBAT delivered with the second-generation system significantly lowers blood pressure in resistant hypertension with stable, intensive background medical therapy, consistent with studies of the first-generation system for electrical activation of the baroreflex, and provides a safety profile comparable to a pacemaker.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of the American Society of Hypertension - Volume 6, Issue 4, July–August 2012, Pages 270–276
نویسندگان
, , , , , , , , ,