Are the enzymatic methods currently being used to measure bronchoalveolar lavage bile salt levels fit for purpose?
چکیده انگلیسی
The enzymatic kits are not sensitive enough for use in situations where bile salt levels are much below 5 μmol/liter, which is the case in BALF. In addition, reports in the literature of levels significantly below 5 μmol/liter need reassessing. Tandem mass spectrometry with a lower limit of detection of 0.01 μmol/liter should be the method of choice.