Commissural dehiscence: A rare and peculiar cause of porcine valve structural deterioration
چکیده انگلیسی

ObjectiveCalcification is the main cause of structural valve deterioration; however, other causes of failure have been identified, and among them, dehiscence of a commissure from the stent has been reported in several models of porcine valves. The aim of this study was to analyze the rate and mode of occurrence of this complication in first- and second-generation porcine bioprosthetic explants.MethodsAmong 586 porcine xenografts explanted and analyzed at the Institute of Pathological Anatomy of the University of Padua, 17 (2.9%) have been replaced for incompetence because of commissural dehiscence. All these explants were in the mitral position, with the exception of a Carpentier-Edwards supra-annular aortic valve prosthesis.ResultsDehiscence was observed in 9 (1.9%) of 455 Hancock standard explants, in 1 (3.2%) of 31 Hancock II, in 3 (8.6%, 2 standard and 1 supra-annular) of 35 Carpentier-Edwards, in 1 (2.4%) of 42 Bioimplants, and in 3 (50%) of 6 Xenotech after a mean time function of 157 ± 50, 156, 96 ± 29, 143, and 130 ± 8 months, respectively. Dehiscence was the sole cause of incompetence in 6 cases. An impending commissural dehiscence caused by blood creeping was observed in one case. This might be an explanation for the dehiscence other than excessive trimming of the aortic wall.ConclusionsCommissural dehiscence is an uncommon and peculiar mode of failure of porcine valves implanted in the mitral position and was observed earlier and more frequently with Carpentier-Edwards porcine explants (P < .05). We speculate that pericardial strip protection of the suture between the Dacron fabric and porcine aortic wall, as used in the Biocor porcine valve, might prevent this complication.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Thoracic and Cardiovascular Surgery - Volume 132, Issue 5, November 2006, Pages 1017–1022
نویسندگان
, , , , , , , ,