Ultrasound-determined diameter measurements are more accurate than axial computed tomography after endovascular aortic aneurysm repair
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Ultrasound-determined diameter measurements are more accurate than axial computed tomography after endovascular aortic aneurysm repair
چکیده انگلیسی

ObjectiveThis study evaluated the correlation of ultrasound (US)-derived aortic aneurysm diameter measurements with centerline, three-dimensional (3-D) reconstruction computed tomography (CT) measurements after endovascular aortic aneurysm repair (EVAR).MethodsConcurrent CT and US examinations from 82 patients undergoing post-EVAR surveillance were reviewed. The aortic aneurysm diameter was defined as the major axis on the centerline images of 3-D CT reconstruction. This was compared with US-derived minor and major axis measurements, as well as with the minor axis measurement on the conventional axial CT images. Correlation was evaluated with linear regression analyses. Agreement between different imaging modalities and measurements was assessed with Bland-Altman plots.ResultsThe correlation coefficients from linear regression analyses were 0.92 between CT centerline major and US minor measurements, 0.94 between CT centerline major and US major measurements, and 0.93 between CT minor and centerline major measurements. Bland-Altman plots showed a mean difference of 0.11 mm between US major and CT centerline measurements compared with 5.38 mm between US minor and CT centerline measurements, and 4.25 mm between axial CT minor and centerline measurements. This suggested that, compared with axial CT and US minor axis measurements, US major axis measurements were in better agreement with CT centerline measurements. Variability between major and minor US and CT centerline diameter measurements was high (standard deviation of difference, 4.27-4.84 mm). However, high variability was also observed between axial CT measurements and centerline CT measurements (standard deviation of difference, 4.36 mm).ConclusionsThe major axis aneurysm diameter measurement obtained by US imaging for surveillance after EVAR correlates well and is in better agreement with centerline 3-D CT reconstruction diameters than axial CT.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Vascular Surgery - Volume 51, Issue 6, June 2010, Pages 1381–1389
نویسندگان
, , , , , ,