ObjectiveTo present our initial experience with the Zenith bifurcated iliac side branch device that preserves internal iliac artery flow whilst excluding aorto-iliac aneurysms.MethodsBetween November 2005 and October 2006, data was prospectively collected on 8 patients in whom this device was used; 2 aorto-bi-iliac aneurysms, 3 aorto-uni-iliac aneurysms, 1 solitary common iliac aneurysm, 1 distal type 1 endoleak, and 1 internal iliac aneurysm.ResultsNo mortality or major complications resulted from use of this device. The median fluoroscopy time was 53 minutes (range 38 to 105) and a median of 102 g of iodine (range 84 to 130) as contrast were used. One patient required a blood transfusion and only one of the eight side branches occluded. There has been no endoleak related to the device in the median follow-up period of 6 months (1 to 14 months).ConclusionThis device provides an alternative for the management of patients with aorto-iliac aneurysms that is safe and less complex than, previously described, hybrid procedures that preserve internal iliac flow.
Journal: Journal of Vascular Surgery - Volume 46, Issue 2, August 2007, Pages 211–217