کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
2995690 1179926 2008 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
First report of a late type III endoleak from fabric tears of a Zenith stent graft
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
First report of a late type III endoleak from fabric tears of a Zenith stent graft
چکیده انگلیسی

We report a case of a late type III endoleak from a hole in the fabric of the main body of a Zenith bifurcated endograft 7 years after implantation. Abdominal pain and a rapidly expanding aneurysm were eventually followed by rupture. The defect was detected at open surgery, whereas no evidence of endoleak was found at preoperative computed tomography (CT) or angiogram. The defect was repaired by a relining procedure with an Excluder stent graft. The patient, however, died 3 weeks after admission.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Vascular Surgery - Volume 48, Issue 3, September 2008, Pages 723–726