Comparison of optic laryngoscope Airtraq® and Miller laryngoscope for tracheal intubation during infant cardiopulmonary resuscitation
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Comparison of optic laryngoscope Airtraq® and Miller laryngoscope for tracheal intubation during infant cardiopulmonary resuscitation
چکیده انگلیسی

BackgroundRecent resuscitation guidelines for infant cardiopulmonary resuscitation (CPR) emphasize that all rescuers should minimize interruption of chest compressions, even for endotracheal intubation. We compared the utility of the Miller laryngoscope (Mil) with Airtraq (ATQ) during chest compression in an infant manikin.MethodsTwenty staff doctors in intensive care and emergency medicine performed tracheal intubation on an infant manikin with Mil and ATQ with or without chest compression.ResultsIn Mil trials, no participants failed without chest compression, but 6 of them failed during chest compression (P < 0.05). In ATQ trials, all participants successfully secured the airway regardless of chest compression. Intubation time was significantly lengthened due to chest compression in Mil trials, but not in ATQ trials. The visual analog scale (VAS) for laryngoscope image did not significantly change due to chest compression for ATQ or Mil trials. In contrast, chest compression worsened VAS scores for tube passage through the glottis in Mil trials, but not in ATQ trials.ConclusionWe conclude that ATQ performed better than Mil for endotracheal intubation during chest compression in infant simulations managed by expert doctors.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Resuscitation - Volume 82, Issue 6, June 2011, Pages 736–739
نویسندگان
, , , , , ,