کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3048669 1185090 2006 11 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Minimization of cochlear implant stimulus artifact in cortical auditory evoked potentials
موضوعات مرتبط
علوم زیستی و بیوفناوری علم عصب شناسی عصب شناسی
پیش نمایش صفحه اول مقاله
Minimization of cochlear implant stimulus artifact in cortical auditory evoked potentials
چکیده انگلیسی

ObjectiveTo compare two methods of minimizing cochlear implant artifact in cortical auditory evoked potential (CAEP) recordings.MethodsTwo experiments were conducted. In the first, we assessed the use of independent component analysis (ICA) as a pre-processing filter. In the second, we explored the use of an optimized differential reference (ODR) for minimizing artifacts.ResultsBoth ICA and the ODR can minimize the artifact and allow measurement of CAEP responses.ConclusionsWhen using a large number of recording electrodes ICA can be used to minimize the implant artifact. When using a single electrode montage an optimized differential reference is adequate to minimize the artifact.SignificanceThe use of an optimized differential reference could allow cortical evoked potentials to be used in routine clinical assessment of auditory pathway development in children and adults fit with cochlear implants.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Neurophysiology - Volume 117, Issue 8, August 2006, Pages 1772–1782