کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3086112 1189846 2009 12 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Amplitude-Integrated Electroencephalography in Neonates
موضوعات مرتبط
علوم زیستی و بیوفناوری علم عصب شناسی علوم اعصاب تکاملی
پیش نمایش صفحه اول مقاله
Amplitude-Integrated Electroencephalography in Neonates
چکیده انگلیسی

Conventional electroencephalography (EEG) has been used for decades in the neonatal intensive care unit for formulating neurologic prognoses, demonstrating brain functional state and degree of maturation, revealing cerebral lesions, and identifying the presence and number of electrographic seizures. However, both the immediate availability of conventional EEG and the expertise with which it is interpreted are variable. Amplitude-integrated EEG provides simplified monitoring of cerebral function, and is rapidly gaining popularity among neonatologists, with growing use in bedside decision making and inclusion criteria for randomized clinical studies. Nonetheless, child neurologists and neurophysiologists remain cautious about relying solely on this tool and prefer interpreting conventional EEG. The present review examines the technical aspects of generating, recording, and interpreting amplitude-integrated EEG and contrasts this approach with conventional EEG. Finally, several proposed amplitude-integrated EEG classification schemes are reviewed. A clear understanding of this emerging technology of measuring brain health in the premature or sick neonate is critical in modern care of the newborn infant.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Pediatric Neurology - Volume 41, Issue 5, November 2009, Pages 315–326